Braun nätverkskort MCA166-USB Manual

PDF-manualer
72 Sidor Engelska

Braun nätverkskort MCA166-USB
MCA166-USB
Name
MMCA_QUERY_POWER
Description The command reads the MMCA power state.
Parameter -
Result array Battery current [mA] long value
Byte Offset 4 HV primary current [mA] long value
Byte Offset 8 +12 primary current [mA] long value
Byte Offset 12 -12 primary current [mA] long value
Byte Offset 16 +24 primary current [mA] long value
Byte Offset 20 -24 primary current [mA] long value
Byte Offset 24 Battery voltage [mV] long value
Byte Offset 28 HV [V]
27
long value
Byte Offset 32 HV state long value
Byte Offset 36
Byte Offset 40 Current high voltage [V]
Byte Offset 44
Byte Offset 48 Power Switches long value 0X80 -24V ON
0X40 +24V ON
0X20 -12V ON
0X10 +12V ON
0X02 HV ON
Byte Offset 52 Charger current [mA] long value
...
Byte Offset 64 Battery current [mA] at stop long value
Byte Offset 68 HV primary current [mA] at stop long value
Byte Offset 72 +12 primary current [mA] at stop long value
Byte Offset 76 -12 primary current [mA] at stop long value
Byte Offset 80 +24 primary current [mA] at stop long value
Byte Offset 84 -24 primary current [mA] at stop long value
Byte Offset 88 Battery voltage [mV] at stop long value
Byte Offset 92 HV [V]
28
at stop long value
Byte Offset 116 Charger current [mA] at stop long value
...
Byte Offset 128 mca_state
29
integer value
Return Code see MMCA_RESET command
27 multiply value with 1.2 V
28 multiply value with 1.2 V
29 see MMCA_QUERY_SYSTEM_DATA
50